Verifying the capacitance function of the 8588A Reference Multimeter

Document(s): 
application/pdf icon Verifying the capacitance function of the 8588A Reference Multimeter (1.21 MB)

Overview: This application note provides guidance to those wishing to verify the performance of the capacitance function of their 8588A Reference Multimeter. It describes the methodology, examines the measurement process, and describes some of the pitfalls that can occur.

As capacitance is derived from resistance, do you need to routinely verify the capacitance function at all? To answer this, first look at how the 8588A Reference Multimeter measures capacitance.

The 8588A measures capacitance using a DC discharge method that is fundamentally the same as the method used by handheld multimeters, but of a higher precision. In the case of the 8588A, the unknown capacitor is discharged until it is close to zero and then charged with a known constant current; this results in a negative going linear ramp that returns to zero at the end of each measurement cycle. The voltages close to the start and end of the ramp are measured and the capacitance calculated from C = I.ẟt/ẟv where C is the capacitance in Farads, I is the current in amps and v is the change in voltage over the time interval t.

The resistance function provides the constant current and voltage measurement capability and uses the same calibration constants for voltage and current that were determined during resistance adjustment, except for the 1 nF range, which is adjusted using an external reference capacitor due to the difficulty in characterizing the internal bias currents.

The capacitance function is also affected by timing errors. Internally the 8588A has one master clock, so these timing errors can be evaluated by measuring the frequency counter. The clock specification is 1 µHz/Hz, which is many orders better than the specification of the best capacitance range, so any timing errors can be effectively ignored.

The Fluke Calibration 8588A Reference Multimeter

The 8588A measures 2 terminal capacitance using the DC discharge method

From this description you can see that the performance of the 10 nF range and above could be monitored by evaluation of the appropriate resistance range. For many laboratories this may be the preferred option due to lack of suitably traceable standards. It is possible that the capacitance measurement could be affected by internal physical changes that would not show as a change in resistance; however, any change would have a greater impact on the lower capacitance ranges, which can be evaluated by comparison with readily available reference capacitors. Any change seen on these lower ranges would very quickly become insignificant as you go up range.

To answer the question; do you need to separately verify the capacitance function? Yes.

For the 1 nF range because its calibration is independent of the resistance range it uses, but optional for the 10 nF range and above. This is because the performance of these ranges can be monitored by observing the change in the appropriate resistance range given in Table 1.

Capacitance range Resistance range used
1 nF 100 MΩ, Lo I
10 nF 10 MΩ, Lo I
100 nF 1 MΩ, Lo I
1 µF 100 kΩ, Lo I
10 µF 10 kΩ
100 µF 1 kΩ
1 mF to 100 mF 100 Ω
1 mF to 100 mF Lo I 1 kΩ

Table 1. Resistance to capacitance range mapping

Using reference capacitors to verify the 8588A

The value of a typical reference capacitor commonly found in calibration laboratories changes with frequency and its configuration as a 2-terminal or 3-terminal device. The 8588A measures 2 terminal capacitance using the DC discharge method described above. We therefore need to know what the effective test frequency of the 8588A is, and the 2-terminal value of the reference capacitor at this frequency.

There is no direct equivalence between the ramped waveform of the 8588A and the frequency of a sine wave. However, a good approximation can be made by equating the slew rate of the ramp generated by the 8588A with the maximum slew rate of a sine wave using the relationship: v/s =2.π.f.Vpk where v/s is the slew rate in volts per second, f is the frequency in Hz and Vpk is the peak voltage. Table 2 lists the equivalent test frequencies determined using this relationship. Note that the frequency not only changes with range but also where the value lies within the range.

8588A range Applied value % of range Equivalent test frequency
1 nF to 1 mF
 
20% 5.63 Hz
100% 1.13 Hz
200% 0.56 Hz
10 mF
 
20% 0.68 Hz
100% 0.14 Hz
200% 0.079 Hz
100 mF
 
20% 0.056 Hz
50% 0.023 Hz
100% 0.011 Hz
1 mF Lo I
 
20% 0.56 Hz
100% 0.11 Hz
200% 0.056 Hz
10 mF Lo I
 
20% 0.056 Hz
100% 0.011 Hz
200% 0.005 6 Hz
100 mF Lo I
 
20% 0.005 6 Hz
50% 0.002 3 Hz
100% 0.001 1 Hz

Table 2. 8588A effective test frequency

It is unlikely that traceability will be available at the frequencies listed in the table, so the reference capacitor will need to be measured at multiple frequencies going as low as possible and the value extrapolated.

Fortunately, the 1 nF range is much simpler to verify. Air dielectric capacitors are available at this value which have negligible frequency dependence; this allows them to be characterized at 1 kHz and used at these low frequencies. The main consideration with these capacitors is that they are generally 3-terminal devices. Therefore, they will need to be modified by fitting a shorting connector to their low terminal that connects it to the case and then measured as a 2-terminal device.

An alternative approach

A slightly different approach is used at Fluke due to the volume of instruments going through the laboratory. We use a characterized 5522A Multi-Product Calibrator that has been characterized using an 8588A transfer standard and the previously described 1 nF capacitor.

Technician Sets Up a Calibration Using the 8588A Reference Multimeter and the 5522A Multi-Product Calibrator

8588A and 5222A work together to create an alternate approach

The capacitance traceability for the transfer standard is derived from voltage, current, and time which has been validated against reference capacitors and a standalone measurement system which is also based on voltage, current, and time. This standalone measurement system has been used since the early 90’s by Fluke and is listed in the UKAS schedule of accreditation.

Keep learning

  • Using precision digital multimeters in place of analog null meters for metrological applications
  • Improved verification of a 55XX Series Multi-Product Calibrator with an 8588A Reference Multimeter
  • Three Surprising Features of the 8588A Reference Multimeter

See related products

  • Bench multimeters
  • 8588A Reference Multimeter
  • 5522A Multi-Product Calibrator

Get help

  • Customer and technical support
  • Speak with a calibration product expert about your equipment needs
  • Home
  • Products
    • New Products
    • Electrical Calibration
      • Electrical Standards
      • Electrical Calibrators
      • Bench Multimeters
      • Electrical Calibration Software
    • RF Calibration
      • RF Reference Sources
      • RF Calibration Accessories
      • RF Calibration Software
    • Data Acquisition and Test Equipment
      • Bench Multimeters
      • Data Acquisition
      • Data Acquisition Software
    • Temperature Calibration
      • ITS-90 Temperature Standards
        • ITS-90 Fixed-point cells
        • Standard Platinum Resistance Thermometers
        • Maintenance Apparatus
        • Liquid Nitrogen Comparison Calibrator
        • Resistance Bridges
        • Standard Resistors
      • Calibration Baths
        • Compact Calibration Baths
        • Standard Calibration Baths
        • Special Application Baths
        • Bath Accessories
        • Bath Controllers
        • Bath Fluids
      • Industrial Calibrators
        • Field Metrology Wells
        • Metrology Wells
        • Handheld Calibrators
        • Field Dry-Block Calibrators
        • Micro Baths
        • Infrared Calibrators
        • Thermocouple Furnaces
        • Dual Block Dry-Well
        • Zero-point Dry-Well
      • Probes / Sensors
        • Platinum Resistance Thermometers (PRTs)
        • Thermistors
        • Thermocouples
      • Digital Thermometer Readouts
      • Multifunction Calibrators
      • Temperature Calibration Software
    • Humidity Calibration
      • Humidity Generators
      • Humidity Data Loggers and Monitors
    • Pressure Calibration
      • Piston Gauges
        • NMI Piston Gauges
        • Absolute Piston Gauges
        • High Pressure Pneumatic Piston Gauges
        • Hydraulic Piston Gauges
        • Piston Gauge Accessories
      • Automated Pressure Controller / Calibrators
        • Low Pressure Controllers / Calibrators
        • Pneumatic Pressure Controllers / Calibrators
        • High Pressure Pneumatic Controller / Calibrators
        • Hydraulic Controller / Calibrators
      • Pressure Monitors
        • Low Pressure Monitors
        • Digital Pressure Gauges
        • Reference Pressure Monitors
      • Deadweight Testers
        • Pneumatic Deadweight Testers
        • Oil Deadweight Testers
        • Water Deadweight Testers
        • High-Pressure Hydraulic Deadweight Testers
        • Deadweight Tester Accessories
      • Manual Pressure Calibration
        • Pressure Calibrators
        • Pneumatic Pressure Control
        • Hydraulic Pressure Comparators / Pumps
      • Handheld Pressure Calibrators
      • Air Data Calibration
      • Environmental Monitors
      • Pressure Calibration Accessories
      • Pressure Calibration Custom Systems
      • Pressure Calibration Software
    • Flow Calibration
      • Gas Flow Standards
      • Gas Flow Accessories
      • GFS Primary Gravimetric Flow Standard
      • Flow Calibration Software
    • Process Calibration Tools
      • Temperature Calibrators
        • Handheld Temperature Calibrators
        • Dry-Block Calibrators and Micro-Baths
        • Precision Digital Thermometers
        • Temperature Probes
        • Infrared Calibrators
        • Hygro Thermometer with Data Logging
      • Pressure Calibrators
        • Digital Pressure Calibrators
        • Handheld Pressure Calibrators
        • Deadweight Testers
        • Precision Digital Pressure Gauges
        • Calibration Hand Pumps
      • Multifunction Calibrators
      • mA Loop Calibrators
      • Process Calibration Software
    • Calibration Software
      • MET/CAL® Software
      • MET/CAL® Support
      • Asset Management Software
      • Temperature Calibration Software
      • Pressure Calibration Software
      • Flow Calibration Software
      • Mechanical / Dimensional Calibration Software
    • Service and Support
    • All Calibration Instruments
  • Purchase Info
    • Where to Buy
    • Request a Quote
    • Request a Demo
    • Request a Sales Consultation
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • News
    • Press Releases
    • Promotions
    • Industry Links
    • Metrology Salary Survey
  • Training and Events
    • Conferences and Exhibitions
    • Training Courses
    • User Group Meetings
    • Web Seminars
      • Live Seminars
      • On-Demand Seminars
  • Literature and Education
    • Education Hub
    • About Calibration
    • Articles and Education
    • Product Literature
    • Product Manuals (User Guides)
    • Videos and Virtual Demos
    • Blog
  • Service and Support
    • Service Request (RMA)
    • Service Plans
    • Technical Support
    • Knowledge Base
    • Accreditations
    • Authorized Service Centers
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • Product Manuals (User Guides)
    • Safety Data Sheets (SDS)
    • Safety, Service, and Product Notices
    • Software Downloads
    • Tools
      • PT100 Calculator
      • ITS-90 Reference Function Calculator
      • Thermocouple Table Voltage Calculator
      • Thermocouple Voltage to Temperature Calculator
  • About Us
    • Contact Us
    • Visit Us
    • Employment
    • Fluke Calibration
      • Hart Scientific Temperature Calibration
      • DH Instruments Pressure and Flow Calibration
      • Pressurements Pressure Calibration
      • Ruska Pressure Calibration
    • Fluke Companies
    • Integrity and Compliance
    • Frequently Asked Questions
Home|Literature|Articles and Education|Electrical Calibration|Application Notes|Verifying the capacitance function of the 8588A Reference Multimeter
©1995-2023 Fluke Corporation
             

Secondary menu

  • Disclaimer
  • Privacy Statement
  • Terms of Use
  • Terms of Sale