Adding Upper and Lower Limits to an EVAL -e Procedure Line
It is sometimes necessary to use an EVAL –e FSC in a procedure that evaluates that a test is within two limits. The following procedure example give several ways to do this evaluation.
=============================================================================
INSTRUMENT: Example of EVAL with upper and lower limits
DATE: 2002-06-03
AUTHOR: William J. Spath
REVISION:
ADJUSTMENT THRESHOLD: 70%
NUMBER OF TESTS: 4
NUMBER OF LINES: 47
=============================================================================
STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON
1.001 LABEL Enter_Number
1.002 MEMI Enter a number between -100 and +100
1.003 JMPL Good_Number ((MEM >= -100) + (MEM <= 100)) > 1
1.004 DISP the number you entered ( [MEM] ) is not between
1.004 DISP -100 and +100!! Try again
1.005 JMPL Enter_Number
1.006 LABEL Good_Number
1.007 MATH M[1] = -100 ; M[2] = -50
1.008 MATH S[1] = "Is " & MEM & " >= " & M[1] & " and <" & M[2]
1.009 EVAL -e ((MEM >= M[1]) + (MEM < M[2])) > 1 : [S1]
2.001 MATH M[1] = -50 ; M[2] = 0
2.002 MATH S[1] = "Is " & MEM & " >= " & M[1] & " and <" & M[2]
2.003 EVAL -e ((MEM >= M[1]) + (MEM < M[2])) > 1: [S1]
3.001 MATH M[1] = 0 ; M[2] = 50
3.002 MATH S[1] = "Is " & MEM & " >= " & M[1] & " and <" & M[2]
3.003 EVAL -e ((MEM >= M[1]) + (MEM < M[2])) > 1: [S1]
4.001 MATH M[1] = 50 ; M[2] = 100
4.002 MATH S[1] = "Is " & MEM & " >= " & M[1] & " and <" & M[2]
4.003 EVAL -e ((MEM >= M[1]) + (MEM < M[2])) > 1: [S1]
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us